Laser Falcon Plus Methane Detector

Caution

The Sensor may be connected/disconnected only when the system is powered off.

  • Attach the Universal Heavy Payload mount to the bottom of your drone.
  • Attach the mounting cover to the back of the sensor.
  • Mount the sensor on the Universal Heavy Payload mount.
  • Connect the power cable to SkyHub power output port #7.
  • Connect the Ethernet data transfer cable to SkyHub Ethernet port #10.

Power on the system.

  • Establish either the drone link or auxiliary connection with the SkyHub in the UgCS-CPM software.
  • On the main screen of the UgCS-CPM, click on the “Payloads” button.
  • From the available payload list, enable the “Laser Falcon Plus methane detector”.
  • Press the “Save” button and confirm system reload.

Wait until the reload process is finished and the connection to the SkyHub is reestablished.

  • On the main screen of the UgCS-CPM, click the “Start” button.
  • From the drop-down widget list, select the “Laser Falcon Plus methane detector” and press the blue “plus” button.
  • To access the configuration menu, press on the “three-dot” icon on the upper right corner of the widget and select “Settings”.

Laser Falcon Plus Methane Detector Configuration Parameters

ParameterDefaultMinMaxPrecisionUnitOptionsDescription
Payload Id2131922201Payload Id
Altitude sourceDefaultLightware SF30D, Ainstein US-D1, Nanoradar NRA24, Mavlink (autopilot), DefaultAltimeter that will be used for altitude data
Minimal starting amplitude3010001mMinimal starting amplitude
Amplitude threshold0010001mAmplitude threshold
Calculation rateNORMALFAST, NORMAL, SLOWCalculation rate.
Concentration threshold002000001mConcentration threshold
Correlation threshold002000001mCorrelation threshold
Data log dir/data/skyhub_logsPath on SkyHub storage where data is saved
Ethernet interfaceEthernet interface to use
Filtered concentration limit10000002000001mFiltered concentration limit
IP address10.2.0.255IP address of the device in the Ethernet network
IP address discovery mode10, 1IP address discovery mode
Leak limit002000001mLeak limit
SNR threshold0010001mSNR threshold
Temperature modeNORMALCOLD, NORMAL, HOTTemperature mode
Action to runNONENONE, REINITIALIZE, FIX_BASEAction to run when logging started
Green laser10, 1, 2, 3, 4Vendor-defined value for green laser pointer mode. Current meaning 0 - Off 1 - On 2 - 0.5Hz 3 - 1Hz 4 - 2Hz Keep in mind that leaving the value 0 (off) is very dangerous since the main laser is invisible IR and there will be no indication that it is activated.