12. FluX Aero Methane Sniffer#

Note

The Sensor may be connected/disconnected only when the system is powered off.

  • Attach the Universal Heavy Payload mount to the bottom of your drone.

  • Attach the gas detector to the Universal Heavy Payload mount.

  • Connect the data cable to the SkyHub Connector #4 (white).

  • Connect the sampling cable to the gas detector through the port with the label “IN”.

  • Attach the sampling arm through the holding clips on the side of the detector.

  • Attach the sampling cable alongside the arm with the provided zip ties.

Power on the system.

  • Establish either the drone link or auxiliary connection with the SkyHub in the UgCS-CPM software.

  • On the main screen of the UgCS-CPM, click on the “Payloads” button.

  • From the available payload list, enable the “FluX Aero Methane Sniffer”.

  • Press the “Save” button and confirm system reload.

Wait until the reload process is finished and the connection to the SkyHub is reestablished.

  • On the main screen of the UgCS-CPM, click the “Start” button.

  • From the drop-down widget list, select the “FluX Aero Methane Sniffer” and press the blue “plus” button.

  • To access the configuration menu, press on the “three-dot” icon on the upper right corner of the widget and select “Settings”.

12.1. FluX methane Gas Detector Configuration Parameters#

Parameter

Default

Min

Max

Precision

Unit

Options

Description

Max concentration

1000

0

10000

1

ppm

Maximal gas concentration to report CPM

Min concentration

0

0

10000

1

ppm

Minimal gas concentration to report in CPM

Analog output Max

1000

0

10000

1

ppm

Maximal gas concentration to report to Analog output

Analog output Min

0

0

10000

1

ppm

Minimal gas concentration to report to Analog output

Analog reference offset

0

0

10000

1

m

Analog reference offset

Analog reference span

0.01

0.0

10000

0.01

m

Analog reference span

Integration time Gas 1

10

0

1000

1

ms

Integration time for gas 1, ms

Gas 1 offset

3

0

10000

1

ms

Concentration gas 1 Offset

Gas 1 span

1.1

0.0

10000

0.01

m

Concentration gas 1 Span

Integration time Gas 2

10

0

1000

1

ms

Integration time for gas 2, ms

Gas 2 offset

1500

0

10000

1

ms

Concentration gas 2 Offset

Gas 2 span

0.95

0.0

10000

0.01

m

Concentration gas 2 Span

Diagnostic mode

False

Diagnostic mode turn on/off

Baud rate

115200

bps

9600, 19200, 115200

Connection port baud rate, 115200 recommended by manufacturer

Mode

Single

Single, Dual

Switch sensor to single/double gases readings

Payload Id

192

192

220

1

Payload Id

Altitude source

Default

Lightware SF30D, Ainstein US-D1, Nanoradar NRA24, Mavlink (autopilot), Default

Altimeter that will be used for altitude data

Primary connector

Connector #4 (AMA3)

Connector #1 (AMA1), Connector #2 (S0), Connector #4 (AMA3)

SkyHub connector